ICJ108 Color Measurements (i1Pro 2)
Measure CIELab values from colors printed on opaque white using ArgyllCMS, average repeated scans, and relate results to ink mix formulas.
Quick start (automated workflow)
./auto-scan.sh
auto-scan.sh runs the whole process in one go:
- Checks required software — verifies
python3and ArgyllCMS (spotread) are available. Installs python3 via apt and downloads ArgyllCMS to~/.local/binautomatically if missing. - Waits for the instrument — prompts you to connect the i1Pro 2 via USB and re-checks until it is detected.
- Scans the samples — lists the sample names from
samples.txtin measurement order, then startsspotread. You first calibrate on the white tile, then take one reading per sample with the instrument button, and press Q when all readings are done. If the number of readings doesn't matchsamples.txt, the scan can be discarded toarchive/and repeated. - Saves a labelled CSV — converts the spotread TSV log to
scan_<mode>_<timestamp>.csvwith asample_idcolumn taken fromsamples.txt. - Updates the averages — re-averages all complete scan CSVs in
the folder into
averaged.csv(L*, a*, b* mean and standard deviation per sample), then asks whether to rebuild the formula tableformula.csv(default: yes).
Options
./auto-scan.sh # D50 Lab (default)
./auto-scan.sh M0 # ISO 13655 M0
./auto-scan.sh M1 # ISO 13655 M1
./auto-scan.sh M2 # ISO 13655 M2 (UV cut)
./auto-scan.sh --average-only # no scanning; just rebuild averaged + formula CSVs
./auto-scan.sh --help # usage
Note: scans made in different modes are kept separate; the averaged file is built only from scans of the mode you pass (default D50).
Prerequisites
- Linux with bash, python3 (auto-installed if missing)
- ArgyllCMS (auto-downloaded to
~/.local/bin/Argyll_V3.4.1if missing) - X-Rite i1Pro 2 with its white calibration tile
- (Recommended, one time) udev rules so the instrument works without root:
sudo ./setup-udev.sh
# then unplug and replug the i1Pro 2
Sample list
samples.txt defines the sample names, one per line, in measurement
order. Lines starting with # are ignored. Mix samples use the pattern
INK_A+INK_B[w1 to w2 mix], e.g.:
ICJ/107
ICJ/108
ICJ/435
ICJ/407
ICJ/107+ICJ/407[75 to 15 mix]
ICJ/107+ICJ/435[30 to 80 mix]
The weights are converted to percentages in the formula table
(e.g. 30 to 80 → 27% / 73%).
Output files
| File | Contents |
|---|---|
scan_D50_<timestamp>.tsv |
Raw spotread log (Lab + XYZ + spectrum 380–730 nm) |
scan_D50_<timestamp>.csv |
Same data as CSV with sample_id column |
averaged.csv |
Per-sample L*, a*, b* mean and std across all scans |
formula.csv |
Averaged Lab plus one percentage column per ink |
archive/ |
Discarded or superseded scans |
Individual tools
These are used by auto-scan.sh but can also be run on their own:
| Script | Purpose |
|---|---|
measure.sh [mode] |
Just the spotread measurement session (TSV output) |
run-measurement.sh [mode] |
Measurement session + CSV conversion |
tsv-to-csv.py <tsv> [-s samples.txt] |
Convert a spotread TSV to labelled CSV |
average-results.py <csv...> -o out.csv |
Average Lab across scan CSVs |
build-formula-table.py |
Rebuild formula.csv from the averaged CSV |
rebuild-average.sh |
Shortcut for ./auto-scan.sh --average-only |
setup-udev.sh |
Install udev rules for USB instrument access (sudo) |
Measurement tips
- Calibrate at the start of each session; recalibrate (press
kin spotread) if the ambient temperature changes. - Keep the instrument flat on the sample and fully covering the aperture.
- Wait ~15 s after calibration before the first reading for best repeatability.
- If repeatability degrades, check for i1Pro lamp drift:
spotread -Y L -O /dev/null, then let the instrument cool a few minutes.