# ICJ108 Color Measurements (i1Pro 2)
|
||
|
||
Measure CIELab values from colors printed on opaque white using ArgyllCMS,
|
||
average repeated scans, and relate results to ink mix formulas.
|
||
|
||
## Quick start (automated workflow)
|
||
|
||
```bash
|
||
./auto-scan.sh
|
||
```
|
||
|
||
`auto-scan.sh` runs the whole process in one go:
|
||
|
||
1. **Checks required software** — verifies `python3` and ArgyllCMS
|
||
(`spotread`) are available. Installs python3 (via `apt` on Linux or
|
||
Homebrew on macOS) and downloads ArgyllCMS to `~/.local/bin`
|
||
automatically if missing.
|
||
2. **Waits for the instrument** — prompts you to connect the i1Pro 2 via
|
||
USB and re-checks until it is detected.
|
||
3. **Scans the samples** — lists the sample names from `samples.txt` in
|
||
measurement order, then starts `spotread`. You first calibrate on the
|
||
white tile, then take one reading per sample with the instrument
|
||
button. After each reading the terminal prints the **next sample
|
||
code** so you know which one to scan next; press **Q** when all
|
||
readings are done. If the number of readings doesn't match
|
||
`samples.txt`, the scan can be discarded to `archive/` and repeated.
|
||
4. **Saves a labelled CSV** — converts the spotread TSV log to
|
||
`scan_<mode>_<timestamp>.csv` with a `sample_id` column taken from
|
||
`samples.txt`.
|
||
5. **Updates the averages** — re-averages **all** complete scan CSVs in
|
||
the folder into `averaged.csv` (L\*, a\*, b\* mean and standard
|
||
deviation per sample, plus a `dE00` CIEDE2000 column against each
|
||
sample's named reference) and a formatted `averaged.html` view, then asks
|
||
whether to rebuild the formula table `formula.csv` (default: yes).
|
||
|
||
### Options
|
||
|
||
```bash
|
||
./auto-scan.sh # D50 Lab (default)
|
||
./auto-scan.sh M0 # ISO 13655 M0
|
||
./auto-scan.sh M1 # ISO 13655 M1
|
||
./auto-scan.sh M2 # ISO 13655 M2 (UV cut)
|
||
./auto-scan.sh --average-only # no scanning; just rebuild averaged + formula CSVs
|
||
./auto-scan.sh --help # usage
|
||
```
|
||
|
||
Note: scans made in different modes are kept separate; the averaged file
|
||
is built only from scans of the mode you pass (default D50).
|
||
|
||
## Prerequisites
|
||
|
||
### Operating system
|
||
|
||
The scripts run on **Linux** and **macOS**:
|
||
|
||
- **Linux** — the primary target. Uses `apt` to install python3 if missing
|
||
and udev rules for USB access (see below). Ships bash 4+.
|
||
- **macOS** — supported. The default `/bin/bash` is 3.2, so `auto-scan.sh`
|
||
is written to stay compatible with it; no newer bash is required. python3
|
||
is installed via [Homebrew](https://brew.sh) if missing, and ArgyllCMS
|
||
uses the macOS build (runs under Rosetta 2 on Apple Silicon). macOS needs
|
||
no udev rules. Note: only the desktop workflow is verified on Linux; the
|
||
macOS paths are portability fixes and have not been hardware-tested.
|
||
- **Windows** — not supported (no Bourne shell / udev; use WSL instead).
|
||
|
||
### Other requirements
|
||
|
||
- bash, python3 (auto-installed if missing)
|
||
- ArgyllCMS (auto-downloaded to `~/.local/bin/Argyll_V3.4.1` if missing)
|
||
- X-Rite i1Pro 2 with its white calibration tile
|
||
- (Linux only, recommended, one time) udev rules so the instrument works
|
||
without root:
|
||
|
||
```bash
|
||
sudo ./setup-udev.sh
|
||
# then unplug and replug the i1Pro 2
|
||
```
|
||
|
||
## Sample list
|
||
|
||
`samples.txt` defines the sample names, one per line, in measurement
|
||
order. This file is **not tracked in git** (it holds your specific
|
||
material codes); copy the template to create your own:
|
||
|
||
```bash
|
||
cp samples.example.txt samples.txt
|
||
```
|
||
|
||
Lines starting with `#` are ignored. Material names can be any codes.
|
||
Append ` > REF` to measure that sample's `dE00` against another sample; a
|
||
bare name (no `>`) is itself a **reference**. Mix samples use the pattern
|
||
`NAME_A+NAME_B[w1 to w2 mix]`, e.g.:
|
||
|
||
```text
|
||
INK_A
|
||
INK_B > INK_A
|
||
INK_A+INK_B[75 to 15 mix] > INK_A
|
||
INK_A+INK_C[30 to 80 mix] > INK_A
|
||
```
|
||
|
||
The weights are converted to percentages in the formula table
|
||
(e.g. `30 to 80` → 27% / 73%), which adds one column per material and
|
||
discovers the material names automatically from your sample list.
|
||
|
||
## Output files
|
||
|
||
| File | Contents |
|
||
|------|----------|
|
||
| `scan_D50_<timestamp>.tsv` | Raw spotread log (Lab + XYZ + spectrum 380–730 nm) |
|
||
| `scan_D50_<timestamp>.csv` | Same data as CSV with `sample_id` column |
|
||
| `averaged.csv` | Per-sample L\*, a\*, b\* mean and std, plus `dE00` vs each sample's reference |
|
||
| `averaged.html` | Formatted view of `averaged.csv` with color swatches |
|
||
| `formula.csv` | Averaged Lab plus one percentage column per ink |
|
||
| `archive/` | Discarded or superseded scans |
|
||
|
||
## Individual tools
|
||
|
||
These are used by `auto-scan.sh` but can also be run on their own:
|
||
|
||
| Script | Purpose |
|
||
|--------|---------|
|
||
| `measure.sh [mode]` | Just the spotread measurement session (TSV output) |
|
||
| `run-measurement.sh [mode]` | Measurement session + CSV conversion |
|
||
| `prompt-next-sample.py -s samples.txt -- spotread …` | Wrap spotread and print the next sample code after each reading |
|
||
| `tsv-to-csv.py <tsv> [-s samples.txt]` | Convert a spotread TSV to labelled CSV |
|
||
| `average-results.py <csv...> -o out.csv` | Average Lab across scan CSVs (writes CSV + HTML) |
|
||
| `build-formula-table.py` | Rebuild `formula.csv` from the averaged CSV |
|
||
| `rebuild-average.sh` | Shortcut for `./auto-scan.sh --average-only` |
|
||
| `setup-udev.sh` | Install udev rules for USB instrument access (sudo) |
|
||
|
||
## Measurement tips
|
||
|
||
- Calibrate at the start of each session; recalibrate (press `k` in
|
||
spotread) if the ambient temperature changes.
|
||
- Keep the instrument flat on the sample and fully covering the aperture.
|
||
- Wait ~15 s after calibration before the first reading for best
|
||
repeatability.
|
||
- If repeatability degrades, check for i1Pro lamp drift:
|
||
`spotread -Y L -O /dev/null`, then let the instrument cool a few
|
||
minutes.
|