Co-authored-by: Cursor <cursoragent@cursor.com>
ICJ108 Color Measurements (i1Pro 2)
Measure CIELab values from colors printed on opaque white using ArgyllCMS, average repeated scans, and relate results to ink mix formulas.
Quick start (automated workflow)
./auto-scan.sh
auto-scan.sh runs the whole process in one go:
- Checks required software — verifies
python3and ArgyllCMS (spotread) are available. Installs python3 (viaapton Linux or Homebrew on macOS) and downloads ArgyllCMS to~/.local/binautomatically if missing. - Waits for the instrument — prompts you to connect the i1Pro 2 via USB and re-checks until it is detected.
- Scans the samples — lists the sample names from
samples.txtin measurement order, then startsspotread. You first calibrate on the white tile, then take one reading per sample with the instrument button. After each reading the terminal prints the next sample code so you know which one to scan next; press Q when all readings are done. If the number of readings doesn't matchsamples.txt, the scan can be discarded toarchive/and repeated. - Saves a labelled CSV — converts the spotread TSV log to
scan_<mode>_<timestamp>.csvwith asample_idcolumn taken fromsamples.txt. - Updates the averages — re-averages all complete scan CSVs in
the folder into
averaged.csv(L*, a*, b* mean and standard deviation per sample, plus adE00CIEDE2000 column against the reference sample) and a formattedaveraged.htmlview, then asks whether to rebuild the formula tableformula.csv(default: yes).
Options
./auto-scan.sh # D50 Lab (default)
./auto-scan.sh M0 # ISO 13655 M0
./auto-scan.sh M1 # ISO 13655 M1
./auto-scan.sh M2 # ISO 13655 M2 (UV cut)
./auto-scan.sh --average-only # no scanning; just rebuild averaged + formula CSVs
./auto-scan.sh --help # usage
Note: scans made in different modes are kept separate; the averaged file is built only from scans of the mode you pass (default D50).
Prerequisites
Operating system
The scripts run on Linux and macOS:
- Linux — the primary target. Uses
aptto install python3 if missing and udev rules for USB access (see below). Ships bash 4+. - macOS — supported. The default
/bin/bashis 3.2, soauto-scan.shis written to stay compatible with it; no newer bash is required. python3 is installed via Homebrew if missing, and ArgyllCMS uses the macOS build (runs under Rosetta 2 on Apple Silicon). macOS needs no udev rules. Note: only the desktop workflow is verified on Linux; the macOS paths are portability fixes and have not been hardware-tested. - Windows — not supported (no Bourne shell / udev; use WSL instead).
Other requirements
- bash, python3 (auto-installed if missing)
- ArgyllCMS (auto-downloaded to
~/.local/bin/Argyll_V3.4.1if missing) - X-Rite i1Pro 2 with its white calibration tile
- (Linux only, recommended, one time) udev rules so the instrument works without root:
sudo ./setup-udev.sh
# then unplug and replug the i1Pro 2
Sample list
samples.txt defines the sample names, one per line, in measurement
order. This file is not tracked in git (it holds your specific
material codes); copy the template to create your own:
cp samples.example.txt samples.txt
Lines starting with # are ignored. Material names can be any codes.
Prefix one sample with * to mark it as the Delta-E reference that the
dE00 column is measured against (if none is marked, the first sample is
used). Mix samples use the pattern NAME_A+NAME_B[w1 to w2 mix], e.g.:
*INK_A
INK_B
INK_A+INK_B[75 to 15 mix]
INK_A+INK_C[30 to 80 mix]
The weights are converted to percentages in the formula table
(e.g. 30 to 80 → 27% / 73%), which adds one column per material and
discovers the material names automatically from your sample list.
Output files
| File | Contents |
|---|---|
scan_D50_<timestamp>.tsv |
Raw spotread log (Lab + XYZ + spectrum 380–730 nm) |
scan_D50_<timestamp>.csv |
Same data as CSV with sample_id column |
averaged.csv |
Per-sample L*, a*, b* mean and std, plus dE00 vs the reference |
averaged.html |
Formatted view of averaged.csv with color swatches |
formula.csv |
Averaged Lab plus one percentage column per ink |
archive/ |
Discarded or superseded scans |
Individual tools
These are used by auto-scan.sh but can also be run on their own:
| Script | Purpose |
|---|---|
measure.sh [mode] |
Just the spotread measurement session (TSV output) |
run-measurement.sh [mode] |
Measurement session + CSV conversion |
prompt-next-sample.py -s samples.txt -- spotread … |
Wrap spotread and print the next sample code after each reading |
tsv-to-csv.py <tsv> [-s samples.txt] |
Convert a spotread TSV to labelled CSV |
average-results.py <csv...> -o out.csv |
Average Lab across scan CSVs (writes CSV + HTML) |
build-formula-table.py |
Rebuild formula.csv from the averaged CSV |
rebuild-average.sh |
Shortcut for ./auto-scan.sh --average-only |
setup-udev.sh |
Install udev rules for USB instrument access (sudo) |
Measurement tips
- Calibrate at the start of each session; recalibrate (press
kin spotread) if the ambient temperature changes. - Keep the instrument flat on the sample and fully covering the aperture.
- Wait ~15 s after calibration before the first reading for best repeatability.
- If repeatability degrades, check for i1Pro lamp drift:
spotread -Y L -O /dev/null, then let the instrument cool a few minutes.