ICJ108 Color Measurements (i1Pro 2)

Measure CIELab values from colors printed on opaque white using ArgyllCMS, average repeated scans, and relate results to ink mix formulas.

Quick start (automated workflow)

./auto-scan.sh

auto-scan.sh runs the whole process in one go:

  1. Checks required software — verifies python3 and ArgyllCMS (spotread) are available. Installs python3 via apt and downloads ArgyllCMS to ~/.local/bin automatically if missing.
  2. Waits for the instrument — prompts you to connect the i1Pro 2 via USB and re-checks until it is detected.
  3. Scans the samples — lists the sample names from samples.txt in measurement order, then starts spotread. You first calibrate on the white tile, then take one reading per sample with the instrument button, and press Q when all readings are done. If the number of readings doesn't match samples.txt, the scan can be discarded to archive/ and repeated.
  4. Saves a labelled CSV — converts the spotread TSV log to scan_<mode>_<timestamp>.csv with a sample_id column taken from samples.txt.
  5. Updates the averages — re-averages all complete scan CSVs in the folder into averaged.csv (L*, a*, b* mean and standard deviation per sample), then asks whether to rebuild the formula table formula.csv (default: yes).

Options

./auto-scan.sh              # D50 Lab (default)
./auto-scan.sh M0           # ISO 13655 M0
./auto-scan.sh M1           # ISO 13655 M1
./auto-scan.sh M2           # ISO 13655 M2 (UV cut)
./auto-scan.sh --average-only   # no scanning; just rebuild averaged + formula CSVs
./auto-scan.sh --help       # usage

Note: scans made in different modes are kept separate; the averaged file is built only from scans of the mode you pass (default D50).

Prerequisites

  • Linux with bash, python3 (auto-installed if missing)
  • ArgyllCMS (auto-downloaded to ~/.local/bin/Argyll_V3.4.1 if missing)
  • X-Rite i1Pro 2 with its white calibration tile
  • (Recommended, one time) udev rules so the instrument works without root:
sudo ./setup-udev.sh
# then unplug and replug the i1Pro 2

Sample list

samples.txt defines the sample names, one per line, in measurement order. Lines starting with # are ignored. Mix samples use the pattern INK_A+INK_B[w1 to w2 mix], e.g.:

ICJ/107
ICJ/108
ICJ/435
ICJ/407
ICJ/107+ICJ/407[75 to 15 mix]
ICJ/107+ICJ/435[30 to 80 mix]

The weights are converted to percentages in the formula table (e.g. 30 to 80 → 27% / 73%).

Output files

File Contents
scan_D50_<timestamp>.tsv Raw spotread log (Lab + XYZ + spectrum 380–730 nm)
scan_D50_<timestamp>.csv Same data as CSV with sample_id column
averaged.csv Per-sample L*, a*, b* mean and std across all scans
formula.csv Averaged Lab plus one percentage column per ink
archive/ Discarded or superseded scans

Individual tools

These are used by auto-scan.sh but can also be run on their own:

Script Purpose
measure.sh [mode] Just the spotread measurement session (TSV output)
run-measurement.sh [mode] Measurement session + CSV conversion
tsv-to-csv.py <tsv> [-s samples.txt] Convert a spotread TSV to labelled CSV
average-results.py <csv...> -o out.csv Average Lab across scan CSVs
build-formula-table.py Rebuild formula.csv from the averaged CSV
rebuild-average.sh Shortcut for ./auto-scan.sh --average-only
setup-udev.sh Install udev rules for USB instrument access (sudo)

Measurement tips

  • Calibrate at the start of each session; recalibrate (press k in spotread) if the ambient temperature changes.
  • Keep the instrument flat on the sample and fully covering the aperture.
  • Wait ~15 s after calibration before the first reading for best repeatability.
  • If repeatability degrades, check for i1Pro lamp drift: spotread -Y L -O /dev/null, then let the instrument cool a few minutes.