# ICJ108 Color Measurements (i1Pro 2)
Measure CIELab values from colors printed on opaque white using ArgyllCMS,
average repeated scans, and relate results to ink mix formulas.
## Quick start (automated workflow)
```bash
./auto-scan.sh
```
`auto-scan.sh` runs the whole process in one go:
1. **Checks required software** — verifies `python3` and ArgyllCMS
(`spotread`) are available. Installs python3 via apt and downloads
ArgyllCMS to `~/.local/bin` automatically if missing.
2. **Waits for the instrument** — prompts you to connect the i1Pro 2 via
USB and re-checks until it is detected.
3. **Scans the samples** — lists the sample names from `samples.txt` in
measurement order, then starts `spotread`. You first calibrate on the
white tile, then take one reading per sample with the instrument
button, and press **Q** when all readings are done. If the number of
readings doesn't match `samples.txt`, the scan can be discarded to
`archive/` and repeated.
4. **Saves a labelled CSV** — converts the spotread TSV log to
`scan_<mode>_<timestamp>.csv` with a `sample_id` column taken from
`samples.txt`.
5. **Updates the averages** — re-averages **all** complete scan CSVs in
the folder into `averaged.csv` (L\*, a\*, b\* mean and standard
deviation per sample), then asks whether to rebuild the formula table
`formula.csv` (default: yes).
### Options
```bash
./auto-scan.sh # D50 Lab (default)
./auto-scan.sh M0 # ISO 13655 M0
./auto-scan.sh M1 # ISO 13655 M1
./auto-scan.sh M2 # ISO 13655 M2 (UV cut)
./auto-scan.sh --average-only # no scanning; just rebuild averaged + formula CSVs
./auto-scan.sh --help # usage
```
Note: scans made in different modes are kept separate; the averaged file
is built only from scans of the mode you pass (default D50).
## Prerequisites
- Linux with bash, python3 (auto-installed if missing)
- ArgyllCMS (auto-downloaded to `~/.local/bin/Argyll_V3.4.1` if missing)
- X-Rite i1Pro 2 with its white calibration tile
- (Recommended, one time) udev rules so the instrument works without root:
```bash
sudo ./setup-udev.sh
# then unplug and replug the i1Pro 2
```
## Sample list
`samples.txt` defines the sample names, one per line, in measurement
order. Lines starting with `#` are ignored. Mix samples use the pattern
`INK_A+INK_B[w1 to w2 mix]`, e.g.:
```text
ICJ/107
ICJ/108
ICJ/435
ICJ/407
ICJ/107+ICJ/407[75 to 15 mix]
ICJ/107+ICJ/435[30 to 80 mix]
```
The weights are converted to percentages in the formula table
(e.g. `30 to 80` → 27% / 73%).
## Output files
| File | Contents |
|------|----------|
| `scan_D50_<timestamp>.tsv` | Raw spotread log (Lab + XYZ + spectrum 380–730 nm) |
| `scan_D50_<timestamp>.csv` | Same data as CSV with `sample_id` column |
| `averaged.csv` | Per-sample L\*, a\*, b\* mean and std across all scans |
| `formula.csv` | Averaged Lab plus one percentage column per ink |
| `archive/` | Discarded or superseded scans |
## Individual tools
These are used by `auto-scan.sh` but can also be run on their own:
| Script | Purpose |
|--------|---------|
| `measure.sh [mode]` | Just the spotread measurement session (TSV output) |
| `run-measurement.sh [mode]` | Measurement session + CSV conversion |
| `tsv-to-csv.py <tsv> [-s samples.txt]` | Convert a spotread TSV to labelled CSV |
| `average-results.py <csv...> -o out.csv` | Average Lab across scan CSVs |
| `build-formula-table.py` | Rebuild `formula.csv` from the averaged CSV |
| `rebuild-average.sh` | Shortcut for `./auto-scan.sh --average-only` |
| `setup-udev.sh` | Install udev rules for USB instrument access (sudo) |
## Measurement tips
- Calibrate at the start of each session; recalibrate (press `k` in
spotread) if the ambient temperature changes.
- Keep the instrument flat on the sample and fully covering the aperture.
- Wait ~15 s after calibration before the first reading for best
repeatability.
- If repeatability degrades, check for i1Pro lamp drift:
`spotread -Y L -O /dev/null`, then let the instrument cool a few
minutes.